To ensure a consistently high quality standard, all sensors and the pressure control system must be inspected and calibrated by the manufacturer after six months of initial operation and annually thereafter.
These services are carried out on-site by the manufacturer or an authorized service provider within Germany and neighboring countries. All required measuring instruments are certified with DKD or PTB calibration certificates and are subject to regular metrological monitoring.